Experiment 1

I. Purpose: In this experiment, data analysis is used and typical lab measures are determined. Samples of various weights are measured and analyzed to determine the accuracy of laboratory tools such as micropipets, pipets, and an analytical balance.

II. Theory: An analytical balance is used throughout this experiment to measure dry beakers and then beakers filled with various amounts of room temperature deionized water. Pipets and micropipettes are used to measure the amount of water. Statistical analysis is then used to determine the accuracy of these tools.

III. Sample Calculations:

Average/Mean: g=30.6860g [pic 1]

Range: Difference of highest and lowest value= 30.6864g-30.6856g=0.0008g

Standard deviation: =0.000356g[pic 2][pic 3]

Relative standard deviation: RSD=== 1.16e-5[pic 4][pic 5]

Temperature adjustment for water: at 23° water is [pic 6]

) = 10.0719mL [pic 7]

Accuracy percentage= [pic 8]

IV. Data and Results:

Mean Mass of Beaker Analysis                               Volume of 10mL Pipet Analysis

30.6863g

30.6864g

30.6859g

30.6856g

30.6857g

Beaker

Dry weight of beaker

Pipet 10 mL into glass

Mass difference

Temperature Adjustment[pic 9]

1

28.9460g

38.9828g

10.0368g

10.0719mL

2

30.6843g

40.6920g

10.0077g

10.0427mL

3

28.6868g

38.4070g

9.7202g

9.7542mL

4

28.8230g

38.8060g

9.9830g

10.0179mL


Volume of 5mL Pipet Analysis[pic 10]

Beaker

Dry weight of beaker

Pipet 5 mL into glass

Mass difference

Temperature Adjustment[pic 11]

1

28.9273g

33.9278g

5.0005g

5.018002mL

2

30.6839g

35.6655g

4.9816g

4.999036mL

3

28.6580g

33.6918g

5.0338g

5.051418mL

4

28.8124g

33.8007g

4.9883g

5.005759mL

Measurements and conversions of 500μL, 1mL (1000μL), 50μL, and 100μL using micropipettes

Note: For the 500μL and 1mL measurements, a 100-1000μL micropipette is used. For the 50μL and 100μL measurements, a 10-100μL micropipette is used.

500μL

Temperature adjustment

1mL (1000 μL)

Temperature adjustment

50μL

Temperature adjustment

100μl

Temperature adjustment

0.5048g

0.5067mL

0.9916g

0.9951mL

0.0497g

0.049874mL

0.1003g

0.1005mL

0.4987g

0.5004mL

0.9951g

0.9986mL

0.0504g

0.05058mL

0.1003g

0.1005mL

0.4978g

0.4995mL

0.9928g

0.9963mL

0.0496g

0.04977mL

0.1004g

0.1006mL

0.4975g

0.4992mL

0.9945g

0.9980mL

0.0499g

0.05008mL

0.1005g

0.1007mL

0.4995g

0.5012mL

0.9922g

0.9957mL

0.0497g

0.04987mL

0.0999g

0.1001mL

...

...